Agilent Pursuit 3 PAH Column Application Note for 24 PAHs Analysis

This document describes a method for the determination of 24 polycyclic aromatic hydrocarbons (PAHs) in drinking water. The analyte is concentrated using an automated solid phase extraction (SPE) method and analyzed by high-performance liquid chromatography (HPLC) with fluorescence detection.


Agilent Pursuit 3 PAH Column Application Note for 24 PAHs Analysis - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
578 KB
Published
11 June, 2012
Updated
29 September, 2023
MD5 checksum
799A8368F7CBC0648DA6F1D37EE949B8

About this manual

The document presents a validated analytical method for the determination of 24 polycyclic aromatic hydrocarbons (PAHs) in drinking water. The method utilizes an automated solid phase extraction (SPE) technique to concentrate analytes from an 800 mL water sample, followed by analysis using High-Performance Liquid Chromatography (HPLC) with fluorescence (FLD) and diode-array (DAD) detectors. The approach achieves a separation of 24 PAHs in approximately 27 to 28 minutes, offering a significant improvement over traditional methods that require around 45 minutes per run.

The analysis relies on the Agilent Pursuit 3 PAH column (100 x 4.6 mm, 3 µm). For the SPE step, the method optimizes the use of the Bond Elut Plexa polymeric sorbent. The process involves loading the sample, drying, and eluting with a mixture of ethyl acetate and dichloromethane. The document details the specific chromatographic conditions, including the gradient profile and fluorescence detection parameters, as well as the validation results demonstrating excellent linearity and precision.

What's inside

  • Introduction
  • Method and Results
  • Chromatographic conditions
  • HPLC gradient profile
  • Sample preparation
  • SPE method
  • Validation Results
  • Conclusion

Specifications

Column dimensions100 x 4.6 mm
Column temperature30 °C
Analysis time27 minutes
SPE Sorbent typeBond Elut Plexa
SPE sample volume800 mL

Frequently asked questions

How long does the HPLC analysis take for 24 PAHs?

The optimized chromatogram led to a separation of 24 PAHs in 27 minutes.

Which sorbent is recommended for the SPE method?

The method achieved best results using the polymer-based sorbent Bond Elut Plexa.