Fast Separation of 27 EU and US EPA Regulated PAHs on Agilent J&W Select PAH Application Note

This application note describes a GC/MS method that resolves all EU and EPA PAHs, and their known interferences, with the Select PAH column in a single run. The method uses enhanced column selectivity and an optimized oven program to resolve all PAHs, including those that co-elute with other PAHs.


Fast Separation of 27 EU and US EPA Regulated PAHs on Agilent J&W Select PAH Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
426 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
23AF98BB400298F910768BEAB7C7C7B8

About this manual

Polycyclic aromatic hydrocarbons (PAHs) are difficult to analyze due to co-elution issues with standard GC/MS methods. This document presents a solution using the Agilent Select PAH column combined with an optimized oven program to separate these compounds effectively.

The note details the specific chromatographic conditions required, including temperature programming and carrier gas flow. It provides data demonstrating the resolution of difficult-to-separate isomers such as benzofluoranthene and dibenzopyrenes, covering both European (EU) and American (EPA 610) regulatory lists.

What's inside

  • Introduction
  • Conditions
  • Results and Discussion
  • Peak Identification
  • Conclusion

Frequently asked questions

What is the runtime of the GC/MS analysis for these PAHs?

The small dimensions of this column delivered a fast runtime of just 29 minutes.

What is the maximum temperature reached during the oven program?

The oven program reaches a final hold at 350 °C for 4 minutes.

How much sample is injected into the system?

The injection volume used in this method is 1.0 pL.