Agilent Dual Column 8081 Pesticide Analysis by GC/ECD with VF-5 Pesticides VF-1701 Pesticide Columns Application Note

This article describes the method and conditions for pesticide analysis by GC/ECD using Agilent J&W FactorFour VF-5 Pesticides and VF-1701 Pesticides columns. The method effectively separates EPA 8081 pesticides and provides peak identification data for VF-1701 Pesticides and VF-5 Pesticides columns.


Agilent Dual Column 8081 Pesticide Analysis by GC/ECD with VF-5 Pesticides VF-1701 Pesticide Columns Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
177 KB
Published
11 June, 2012
Updated
29 September, 2023
MD5 checksum
7247C07F3787A012A47F3C719771469B

About this manual

This document is an application note from Agilent Technologies detailing the analysis of EPA 8081 pesticide standards using Gas Chromatography with Electron Capture Detection (GC-ECD). It specifically evaluates the performance of two new generation FactorFour columns: the VF-5 Pesticides and the VF-1701 Pesticides.

What's inside

  • Conditions
  • Figure 1. Separation of EPA 8081 Pesticides using VF-5
  • Figure 2. Separation of EPA 8081 Pesticides using VF-1701
  • Table 1. Compound list

Specifications

TechniqueGC-ECD
Column TypeFused silica
Column Dimensions30 m x 0.25 mm df = 0.25 ym
Part Number VF-5CP9070
Part Number VF-1701CP9074
Injector Temperature250 °C
Detector Temperature275 °C
Sample Size0.5 yL
Sample Concentration200 ng/mL

Frequently asked questions

Which columns are used for EPA 8081 pesticide analysis in this note?

The analysis uses Agilent J&W FactorFour VF-5 Pesticides and VF-1701 Pesticides columns.

What is the temperature program for the GC analysis?

The temperature starts at 59 °C (hold 30 sec), ramps to 150 °C @ 15 °C per min, then to 275 °C @ 4°C per min.