Agilent 1260 Infinity and 6520 Q-TOF LC/MS Application Note: Identification of Oxidation Sites on a Monoclonal Antibody

This application note describes a method for identifying oxidation sites on a monoclonal antibody (mAb) using an Agilent 1260 Infinity HPLC-Chip/MS system coupled to an Agilent Accurate-Mass 6520 Quadrupole Time-of-Flight (Q-TOF) LC/MS. A monoclonal antibody was oxidized by treatment with hydrogen peroxide to stimulate potential oxidative modifications that may occur during the manufacturing process. The accurate mass measurement of the 6520 Q-TOF LC/MS combined with Agilent MassHunter BioConfirm software facilitated easy identification of the oxidation sites. Peptides containing amino acid residues such as methionine and tryptophan, which are sensitive to oxidation, were easily identified using this approach.


Agilent 1260 Infinity and 6520 Q-TOF LC/MS Application Note: Identification of Oxidation Sites on a Monoclonal Antibody - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
337 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
8C92DD13A695BF35DAD9E33336F9A6DA

About this manual

Monoclonal antibodies can undergo oxidation during production and storage, altering their activity. This document details a method using Agilent's microfluidic-based HPLC-Chip integrated with the Accurate-Mass 6520 Q-TOF LC/MS to analyze potential oxidative modifications in monoclonal antibodies treated with hydrogen peroxide.

The study involved reducing and alkylating the antibody, followed by trypsin digestion. The resulting peptides were analyzed using MassHunter BioConfirm software. Results confirmed that peptides containing methionine and tryptophan residues were readily oxidized, allowing for the identification of specific modification sites through accurate mass measurement.

What's inside

  • Introduction
  • Sample Preparation
  • Oxidation of mAb using hydrogen peroxide
  • Reduction and alkylation of an antibody
  • Trypsin digestion
  • LC/MS Analysis
  • Instrumentation
  • LC Parameters
  • MS Parameters
  • Data Analysis

Specifications

Enrichment Column Dimensions40 nL
Analytical Column Dimensions75 mm x 43 mm
Flow Rate (Nanoflow LC Pump)600 nL/min
Sample Load Amount50 ng
MS Acquisition Range (MS only)300-3,200 m/z
Fragmentor Voltage175 V

Frequently asked questions

Which amino acid residues are sensitive to oxidation in this study?

Peptides containing methionine and tryptophan residues were easily identified as being sensitive to oxidation.

What software was used for data analysis?

The data were analyzed using Agilent MassHunter Qualitative Analysis software and Agilent MassHunter BioConfirm software.

How was the oxidation reaction performed?

A 2 uL sample of antibody (1 mg/mL) was diluted with water, and 2 uL of H2O2 (50% concentration) was added for a 20-minute reaction at room temperature.

Download PDF (337 KB)