Key facts
- Brand
- Rigol
- Model
- T2R1000
- Category
- Adapter
- Industry
- Electronics
- Documents
- 2
- Languages
- ZH
- Last added
- 2026-11-16
Start here

RIGOL T2R1000 Active Probe Adapter User's Guide
The T2R1000 Active Probe Adapter converts a Tektronix TekProbe‑BNC level‑II probe to the RIGOL‑Probe interface, supplying power, calibration and DC‑offset adjustment. Connect the probe to the adapter, then plug the adapter into the analog input of a compatible RIGOL oscilloscope and select the probe model in the oscilloscope menu.
The adapter supports a bandwidth greater than 4 GHz, operates from ±5 V/±15 V supplies, and delivers up to 150 mA output current. It tolerates a maximum input voltage of 42 Vpk (30 Vrms) and works within 0 °C‑50 °C (operating) and –40 °C‑70 °C (non‑operating) temperature ranges. Physical dimensions are 60 × 33.7 × 29.5 mm and weight is 132 g with packaging. This manual is written in Chinese.
RIGOL T2R1000 specifications
| Bandwidth | >4GHz (T2R1000 only) |
|---|---|
| DC Offset Range | <±1V |
| Max Input Voltage | 42Vpk, 30Vrms |
| Temperature | Operation: 0°C-50°C; Non-operation: -40°C-70°C |
| Humidity | 95%RH at 50°C |
| Size | 60mm x 33.7mm x 29.5mm |
| Weight | 132g with package; 41g without package |
All RIGOL T2R1000 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
Which oscilloscopes can be used with the T2R1000?
It can be used with RIGOL DS6000 or MSO4000/DS4000 series digital oscilloscopes, with the required software version or above.
Which probes does the T2R1000 support?
It supports TekProbe-BNC level II probes, including the listed Tektronix active, differential, current, and optical probes.
How do I connect a TekProbe‑BNC probe to a RIGOL oscilloscope using the T2R1000?
Insert the TekProbe‑BNC probe into the T2R1000 adapter, then plug the adapter’s other end into the oscilloscope’s analog input that has a RIGOL‑Probe connector, and select the probe model in the oscilloscope menu.
What is the maximum input voltage the T2R1000 can tolerate?
It can handle up to 42 V peak (30 Vrms).