RIGOL T2R1000 Active Probe Adaptor User Manual

User manual for the RIGOL T2R1000 active probe adaptor connecting TekProbe-BNC probes to RIGOL oscilloscopes.


RIGOL T2R1000 Active Probe Adaptor User Manual - cover page
Brand
Rigol
Category
Adapter
Document type
User Manual
Language
Chinese
Pages
13
File format
PDF
File size
282 KB
Published
16 November, 2026
Updated
16 November, 2026
Model
T2R1000
MD5 checksum
91CF86FED6EC5A7CFDB63624401EDAA2

About this manual

The T2R1000 active probe adaptor connects TekProbe-BNC level II probes to RIGOL DS6000 or MSO4000/DS4000 series digital oscilloscopes. It converts the TekProbe-BNC interface to the RIGOL-Probe interface and provides power, calibration, and DC offset adjustment functions for supported probes.

The manual covers connection methods, oscilloscope menu operation, supported Tektronix active, differential, current, and optical probes, required oscilloscope software versions, and product specifications. This manual is written in Chinese.

What's inside

  • Product Overview
  • Connection Method
  • Menu Operation
  • Tektronix Active Probes Supported by T2R1000
  • RIGOL Oscilloscope Supported by T2R1000
  • Specifications
  • Contact Us

Specifications

Bandwidth>4GHz (T2R1000 only)
DC Offset Range<±1V
Max Input Voltage42Vpk, 30Vrms
TemperatureOperation: 0°C-50°C; Non-operation: -40°C-70°C
Humidity95%RH at 50°C
Size60mm x 33.7mm x 29.5mm
Weight132g with package; 41g without package

Frequently asked questions

Which oscilloscopes can be used with the T2R1000?

It can be used with RIGOL DS6000 or MSO4000/DS4000 series digital oscilloscopes, with the required software version or above.

Which probes does the T2R1000 support?

It supports TekProbe-BNC level II probes, including the listed Tektronix active, differential, current, and optical probes.

How is probe calibration performed?

Connect the adaptor and probe, select the probe model, press Probe-Cal, ground the probe input terminal when prompted, then press OK.

What is the Bias Voltage function used for?

It adjusts a signal under test that exceeds the probe amplifier input dynamic range to a proper range.