Key facts
- Brand
- Rigol
- Model
- RP1050H
- Category
- Probes
- Industry
- Electronics
- Documents
- 2
- Languages
- ZH · EN
- Last added
- 2026-11-18
Start here

RIGOL RP1050H High Voltage Probe User's Guide
This document is the User's Guide for the RIGOL RP1050H High Voltage Probe. It provides critical safety warnings regarding high voltage measurements, including grounding requirements and maximum exposure times for specific voltage levels to prevent probe damage or injury. The guide also outlines proper cleaning procedures and emphasizes that the device must only be operated by trained personnel.
Read onlineRIGOL RP1050H specifications
| Max. Working Voltage DC | 0 ~ 10 KV |
|---|---|
| Max. Working Voltage AC Pulse | ≤ 20 KV peak to peak |
| Max. Working Voltage AC Sine | ≤ 7KV rms |
| Signal/Noise | DC ≥ 60 dB (1 KHz), ≥ 50 dB (1 MHz) |
| Bandwidth | DC ~40 MHz (-3 dB) |
| Division Ratio | 1:1000 |
| Rise Time | ≤ 8.8 ns |
| Input Impedance | 100 MΩ ± 5% |
| Input Capacitance | 3.0 pF ± 0.5 pF |
| Compensation Range | 5 pF ~ 50 pF |
| Operating Temperature | 0 ~ 50℃ |
| Storage Temperature | -20 ~ +70℃ |
Source: RIGOL RP1050H High Voltage Oscilloscope Probe User's Guide
All RIGOL RP1050H documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the maximum DC working voltage?
The maximum DC working voltage is 0 ~ 10 KV.
From: RIGOL RP1050H High Voltage Oscilloscope Probe User's Guide
How long can each high voltage measurement last?
Each measurement should be within 60 seconds, with at least a 5-minute break for cooling.
From: RIGOL RP1050H High Voltage Oscilloscope Probe User's Guide
What is the maximum measurement time for high voltage?
Measurements should not exceed 60 seconds to prevent probe overheating or damage.
Can I measure while my body is touching the device?
No, you must ensure no part of your body is in contact with the device before turning the high voltage on.