Instruction manual for the Hanna Instruments HI 9147 Oxycheck portable dissolved oxygen meter, covering setup, calibration, measurement and maintenance.
The HI 9147 is a waterproof portable dissolved oxygen meter with manual calibration, automatic temperature compensation, salinity compensation and a fixed galvanic probe. Dissolved oxygen readings are shown in mg/L (ppm) or % saturation, and temperature is measured over -5 to 50°C.
The manual explains preliminary checks, controls, specifications, supplied accessories, probe preparation, one-point calibration, measurement, salinity and altitude compensation, probe/membrane maintenance and battery replacement. Two versions are listed: HI 9147-04 with the HI 76409/4 probe (4 m cable) and HI 9147-10 with the HI 76409/10 probe (10 m cable). This manual is written in Portuguese.
| Dissolved oxygen range | 0.0 to 50.0 mg/L (ppm) O2 |
|---|---|
| Oxygen saturation range | 0 to 600% O2 |
| Temperature range | -5.0 to 50.0°C |
| Resolution | 0.1 mg/L or 1% O2; 0.1°C (1°F) |
| Accuracy | ±1% of reading (O2); ±0.2°C (1°F), excluding probe error |
| Calibration | Manual, in saturated air |
| Temperature compensation | Automatic, 0 to 50°C |
| Salinity compensation | 0 to 51 g/L (resolution 1 g/L) |
| Probe | Fixed HI 76409/4 with 4 m cable or HI 76409/10 with 10 m cable |
| Battery | 3 x 1.5V AAA |
| Battery life | 1000 hours of use (backlight off) |
| Auto-off | After approximately 8 minutes |
| Dimensions | 185 x 72 x 36 mm |
| Weight | 450 g with HI 76409/4 probe |
Calibration is manual in saturated air and is a simple one-point procedure.
The manual recommends replacing the membrane every 2 months and the electrolyte once a month.
A flashing “BATT” symbol appears on the LCD to indicate that about 1 hour of operation remains, and the batteries should be replaced as soon as possible.
No. The manual states that accurate readings are not possible in still water.