Hanna Instruments HI9147 Portable Underwater Dissolved Oxygen Meter User Manual

This document is the instruction manual of HI9147 portable underwater dissolved oxygen meter


Hanna Instruments HI9147 Portable Underwater Dissolved Oxygen Meter User Manual - cover page
Brand
Hanna Instruments
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
421 KB
Published
26 October, 2023
Updated
26 October, 2023
Model
HI 9147
MD5 checksum
ED6DA6A74C332622023E7FB47C7D4D7F

About this manual

The HI9147 is a water-resistant dissolved oxygen meter designed for field and lab use. It features automatic temperature compensation, salinity compensation, and manual calibration. Measurements are displayed in mg/L or % saturation, with temperature readings from -5 to 50°C. The meter includes a fixed galvanic probe with a 4m or 10m cable, depending on model variant. It is powered by three AAA batteries and has a rugged, portable design for durability in harsh environments.

The manual covers probe preparation, calibration procedures, measurement techniques, battery replacement, salinity and altitude compensation, and maintenance recommendations. It also includes warranty information and troubleshooting guidance for users.

What's inside

  • Functional Description
  • Probe Preparation
  • Calibration
  • Taking Measurements
  • Battery Replacement
  • Salinity Compensation
  • Altitude Compensation
  • Probe & Membrane Maintenance
  • Warranty
  • Accessories

Specifications

Range0.0 to 50.0 mg/L O2, 0 to 600 % O2, -5.0 to 50.0 °C
Resolution0.1 mg/L or 1% (O2) / 0.1 °C (1 °F)
Accuracy±1% of reading (O2) / ±0.2°C (1°F) (excluding probe error)
CalibrationManual, in saturated air
Temperature CompensationAutomatic, from 0 to 50 °C (32 to 122 °F)
Salinity Compensation0 to 51 g/L (resolution 1 g/L)
ProbeHI76409/4 with 4 m cable or HI76409/10 with 10 m cable
Battery Type3 x 1.5V AAA
Battery Life1000 hours of use (BL off)
Dimensions185 x 72 x 36 mm (7.3 x 2.8 x 1.4")
Weight450 g (15.9 oz.)

Frequently asked questions

How do I calibrate the HI9147?

Calibration is a simple 1-point procedure performed in air. Switch the meter on, select % mode, and adjust the % trimmer to display 100%.

What is the temperature range for the HI9147?

The temperature range is -5.0 to 50.0 °C (23 to 122 °F).

What should I do if the battery level is low?

A blinking “BATT” tag will appear on the LCD. Replace the batteries as soon as possible to avoid interruption of use.

Download PDF (421 KB)

More Hanna Instruments Manuals


View all Hanna Instruments manuals →