Operation manual for the JEOL JEOL-7100F FE-SEM covering sample mounting, observation, image saving, backscattered electron imaging, EDX analysis, and shutdown.
JEOL JEOL-7100F FE-SEM Operation Manual is a Japanese-language operating guide for the JEOL-7100F field emission scanning electron microscope. It describes precautions for use, sample mounting, observation, image saving, backscattered electron image settings, EDX analysis, and sample exchange and shutdown procedures.
The manual also states operating conditions and cautions, including weekday use hours, the vacuum level required before observation, accelerating voltage selection, and EDX analysis settings. This manual is written in Japanese.
| Observation vacuum level | 3.0E-3 Pa or lower |
|---|---|
| Accelerating voltage selection | 15 kV or 5 kV |
| Accelerating voltage change interval | 15 minutes or more |
| EDX accelerating voltage | 15 kV |
| EDX irradiation current | DT not exceeding 25% |
| EDX working distance (Z axis) | 10 |
Stop operation immediately and contact the administrator.
The instrument is available on weekdays from 9:00 to 17:00.
Select either 15 kV or 5 kV.
Wait until the sample chamber vacuum reaches 3.0E-3 Pa or lower.