JEOL JEOL-7100F FE-SEM Operation Manual

Operation manual for the JEOL JEOL-7100F FE-SEM covering sample mounting, observation, image saving, backscattered electron imaging, EDX analysis, and shutdown.


JEOL JEOL-7100F FE-SEM Operation Manual - cover page
Brand
JEOL
Category
microscope
Document type
User Manual
Language
Japanese
Pages
12
File format
PDF
File size
1.5 MB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
93761F6A305D3010A9AD572F8F2D5DDD

About this manual

JEOL JEOL-7100F FE-SEM Operation Manual is a Japanese-language operating guide for the JEOL-7100F field emission scanning electron microscope. It describes precautions for use, sample mounting, observation, image saving, backscattered electron image settings, EDX analysis, and sample exchange and shutdown procedures.

The manual also states operating conditions and cautions, including weekday use hours, the vacuum level required before observation, accelerating voltage selection, and EDX analysis settings. This manual is written in Japanese.

What's inside

  • Operation precautions
  • Sample mounting
  • Observation
  • Image saving
  • Backscattered electron image settings
  • EDX analysis
  • Sample exchange and shutdown

Specifications

Observation vacuum level3.0E-3 Pa or lower
Accelerating voltage selection15 kV or 5 kV
Accelerating voltage change interval15 minutes or more
EDX accelerating voltage15 kV
EDX irradiation currentDT not exceeding 25%
EDX working distance (Z axis)10

Frequently asked questions

What should I do if an abnormality occurs?

Stop operation immediately and contact the administrator.

What are the operating hours?

The instrument is available on weekdays from 9:00 to 17:00.

Which accelerating voltages should be selected?

Select either 15 kV or 5 kV.

What vacuum level is required before observation?

Wait until the sample chamber vacuum reaches 3.0E-3 Pa or lower.

Download PDF (1.5 MB)