User instructions for the JEOL 6700 scanning electron microscope covering specimen exchange, imaging setup, alignment, image acquisition, and frame integration.
This JEOL 6700 user manual describes the operating procedure for the scanning electron microscope. It covers starting conditions, specimen loading and unloading, software setup, obtaining an image, alignment, image acquisition, recording stage positions, and frame integration.
The document is a step-by-step guide to console and software actions, including vacuum checks, specimen holder handling, imaging parameters, alignment adjustments, and saving images to the jsm6700 user files directory.
| Operating vacuum | 10-4 -- 10-5 Pa |
|---|---|
| Specimen exchange vacuum | 5.00E-004 Pa or lower |
The Penning Gauge should show 10-4 to 10-5 Pa; if not, consult with staff.
Vent the chamber, mount the specimen holder with the bottom groove parallel to the loading direction, evacuate, then insert and withdraw the rod as described.
Select the Image File Handler button, choose the jsm6700 user files directory, and select Export to get a micron marker.
Click Stage Map, then Points Map, then Points File; use Record for specific positions and Go to return to them.