Bio-Rad CFX Opus 96, 384, and Deepwell Real-Time PCR Systems Instrument Guide

This user guide covers safety, setup, configuration, protocol creation, running protocols, file management, maintenance, and troubleshooting for Bio-Rad CFX…


Bio-Rad CFX Opus 96, 384, and Deepwell Real-Time PCR Systems Instrument Guide - cover page
Brand
BIO-RAD
Category
Other Electronics
Document type
User Manual
Language
English
Pages
176
File format
PDF
File size
3.7 MB
Published
13 October, 2026
Updated
13 October, 2026
MD5 checksum
936169CE9F9E1E4D92276CAB7290A732

About this manual

This document serves as the Instrument Guide for the CFX Opus 96, CFX Opus 384, and CFX Opus Deepwell Real-Time PCR Systems. It provides comprehensive instructions for researchers on how to operate these instruments for research use only. The guide details the system's main features and performance specifications, including sample block and optical detection capabilities.

What's inside

  • Safety and Regulatory Compliance
  • System Overview
  • Setting Up the CFX Opus System
  • Configuring the CFX Opus System
  • Creating Protocols
  • Running Protocols
  • Managing Files and Folders
  • Maintenance and Troubleshooting

Specifications

Sample Volume (CFX Opus 96)1–50µl (10–50µl Recommended)
Sample Volume (CFX Opus 384)1–30µl (5–30µl Recommended)
Sample Volume (CFX Opus Deepwell)1–125µl (10–125µl Recommended)
Maximum Ramp Rate (CFX Opus 96)5°C/sec
Maximum Ramp Rate (CFX Opus 384/Deepwell)2.5°C/sec
Temperature Range4–100°C
Lid Temperature30–110°C
Temperature Accuracy±0.2°C of programmed target at 90°C
Temperature Uniformity±0.3°C well-to-well within 10sec of arrival at 90°C
Optical Detection Range450-730nm (96/Deepwell), 450-690nm (384)
Multiplex Analysis5 targets per well (96/Deepwell), 4 targets per well (384)
SensitivityDetects 1 copy of target sequence in human genomic DNA

Frequently asked questions

What is the recommended sample volume for the CFX Opus 96 system?

The recommended sample volume is 10–50µl, with a maximum capacity of 50µl.

What is the maximum ramp rate for the CFX Opus 96 system?

The maximum ramp rate for the CFX Opus 96 system is 5°C/sec.

What is the temperature accuracy of the CFX Opus Real-Time PCR Systems?

The temperature accuracy is ±0.2°C of the programmed target at 90°C.

How many targets per well can the CFX Opus 384 system analyze?

The CFX Opus 384 system can perform multiplex analysis for 4 targets per well.

Download PDF (3.7 MB)