This manual introduces the use of Promega's luciferase detection system, which can be used for quantitative determination of luciferase in mammalian cells, as well as in plant and bacterial cells.
The document is a Chinese-language operation manual (CTB281) for the Promega Luciferase Assay System and refers to the English technical manual TB281. It states that it applies to products E1483, E1500, E1501, E4030, E4530 and E4550.
The manual describes assay reagent preparation, three lysis buffers (CCLR, RLB and PLB), preparation of cell lysates, and luciferase measurement using manual luminometers, single-tube luminometers with injectors, and plate readers. It also includes sections on light detection linear range, general precautions, light intensity optimization, instrument use, firefly luciferase reporter vectors, solution and buffer compositions, related products, references, and a protocol for experienced users. This manual is written in Chinese.
| Catalog numbers covered | E1483, E1500, E1501, E4030, E4530, and E4550 |
|---|---|
| Standard assays per E1500/E4030 system | 100 |
| Standard assays per E1501/E4530/E4550 system | 1,000 |
| Luciferase assay reagent per reaction | 100 µl |
| Firefly luciferase molecular weight | 61 kDa |
| Prepared assay reagent stability at -20°C | 1 month |
| Prepared assay reagent stability at -70°C | 1 year |
| Light signal stability | At least 60 seconds |
It covers E1483, E1500, E1501, E4030, E4530, and E4550.
The E1500 system contains enough reagent for 100 standard assays.
It is stable for one year at -70°C.
Each reaction requires 100 µl of luciferase assay reagent.