Agilent ZORBAX StableBond SB-C18 LC Column Application Note for Oxycodone Analysis

This application note describes a method for analyzing oxycodone and its metabolites (noroxycodone, oxymorphone, and noroxymorphone) in plasma using high-performance liquid chromatography/mass spectrometry (HPLC/MS) and the Agilent ZORBAX Rapid Resolution High Throughput (RRHT) StableBond SB-C18 column for chromatographic separation. The method employs an ammonium acetate/acetonitrile gradient and detection by electrospray positive ionization mode mass spectrometry. Spiked human plasma samples are subjected to solid-phase extraction prior to HPLC/MS analysis. The method provides fast, efficient analysis with minimal solvent consumption, which increases productivity, and good linearity (R2 > 0.9900) and reproducibility (less than 10% difference between replicate analyses) for all compounds.


Agilent ZORBAX StableBond SB-C18 LC Column Application Note for Oxycodone Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
Japanese
Pages
6
File format
PDF
File size
190 KB
Published
06 July, 2012
Updated
01 October, 2023
MD5 checksum
9F9A712CF79D09E5C95F903E7190D51C

About this manual

This document details a method for analyzing oxycodone and its oxidation metabolites (noroxycodone, oxymorphone, noroxymorphone) in human plasma. The method combines high-speed liquid chromatography with mass spectrometry detection using an Agilent ZORBAX Rapid Resolution High Throughput StableBond SB-C18 column.

The procedure involves solid-phase extraction of spiked human plasma samples followed by analysis via LC/MS in electrospray positive ion mode. The method achieves good linearity (R² > 0.9900) and reproducibility, with a limit of detection for oxycodone at 0.5 ng/mL. This manual is written in Japanese.

What's inside

  • Introduction
  • Experiment Setup
  • Matrix Sample Preparation
  • Results and Discussion
  • Chromatograms

Frequently asked questions

What is the minimum detection limit for oxycodone using this method?

The detection and quantification limit for oxycodone is 0.5 ng/mL.

What are the dimensions of the ZORBAX StableBond SB-C18 column used?

The column measures 2.1 mm in inner diameter and 50 mm in length.

What is the maximum acceptable difference between replicate samples for reproducibility?

The difference between sets of replicate samples must be less than 10%.

Download PDF (190 KB)