Agilent 6410A Triple Quadrupole Mass Spectrometer Application Note

This document describes the method of analyzing pharmaceuticals in water samples using the Agilent G6410A triple quadrupole mass spectrometer (QQQ) with HPLC/MS/MS. The method was developed following the guidelines of EPA method 1694. Four different chromatographic gradients and LC conditions were used to accommodate the polarity and systematics of various pharmaceuticals. Additionally, analysis using positive and negative ion electrospray was performed for quantification and qualification of each compound.


Agilent 6410A Triple Quadrupole Mass Spectrometer Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
Japanese
Pages
12
File format
PDF
File size
584 KB
Published
06 July, 2012
Updated
29 September, 2023
Model
6410A
MD5 checksum
7F66B883927F74A49585F8C2B4D1EBF3

About this manual

This document details an analytical method developed using the Agilent 6410A Triple Quadrupole Mass Spectrometer (QQQ) to screen and confirm 65 pharmaceuticals in water samples, following EPA Method 1694 guidelines. The method utilizes four different chromatographic gradients and LC conditions based on the polarity and class of the pharmaceuticals, employing both positive and negative ion electrospray ionization (ESI). This manual is written in Japanese.

What's inside

  • Introduction
  • Sample Preparation
  • LC/MS/MS Instrument Settings
  • Optimization of LC/MS/MS Conditions
  • Results and Discussion
  • Analytical Performance

Specifications

Instrument ModelAgilent 6410A Triple Quadrupole Mass Spectrometer
Ionization ModePositive and Negative ESI
Nebulizer Pressure40 psig
Capillary Voltage4000V
Drying Gas Temperature300 C
Fragmenter Voltage70-130V
Collision Energy5-35 V
Dwell Time10 msec

Frequently asked questions

Which EPA method is this analysis based on?

The analysis follows the guidelines of EPA Method 1694.

How many MRM transitions are used for each compound?

Two MRM transitions are used for each compound for quantification and qualitative ion analysis.