Agilent 500 Ion Trap LC/MS System Fast Polarity Switching Technical Overview

Agilent 500 Ion Trap LC/MS System can quickly switch ion polarity to realize real-time positive and negative ion detection


Agilent 500 Ion Trap LC/MS System Fast Polarity Switching Technical Overview - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
4
File format
PDF
File size
404 KB
Published
05 July, 2012
Updated
29 September, 2023
MD5 checksum
08CF80AA046A1F8A0916E3703AB1B5C6

About this manual

This document provides a technical overview of the fast ion polarity switching feature available on the Agilent 500 Ion Trap LC/MS System. The system is capable of changing polarity every 200 milliseconds, which allows for the observation of both positive and negative ions in real time. This speed is significantly faster than many competitive systems that require over one second to switch, making them less effective for real-time dual-polarity detection.

What's inside

  • Technical Overview
  • Product Description
  • Discussion and Results
  • Benefits

Frequently asked questions

How often does the Agilent 500 Ion Trap LC/MS System change polarity?

The system changes polarity every 200 msec.

What is the maximum voltage the electrospray needle can switch between?

The needle voltage can switch from +6 kV to -6 kV.

How long does it take for the conversion dynode voltage to change from -15 kV to +15 kV?

It takes 200 msec for the conversion dynode voltage to change from -15 kV to +15 kV.