Agilent Turbo V80 Pumps vs V81-T Technical Memo Manual

This document is about the technical comparison of V80 series pumps and Turbo 81-T series pumps


Agilent Turbo V80 Pumps vs V81-T Technical Memo Manual - cover page
Brand
Agilent
Category
Other Electronics
Document type
Product Leaflet
Language
English
Pages
13
File format
PDF
File size
388 KB
Published
05 July, 2012
Updated
30 September, 2023
Model
V81-T
MD5 checksum
C83324282275783C5CC0B1867CCFEC20

About this manual

This document serves as a technical memo comparing the Agilent V80 series pumps with the newer Turbo 81-T series pumps. It provides outline drawings for both ISO63 and CFF 4-4 configurations, highlighting dimensional differences and suggesting replacements for the upgrade program. The text notes that while the inlet screen remains the same, other accessories must be replaced during the upgrade.

What's inside

  • Outline drawings
  • Technical Specification
  • Technical Advantages
  • Accessories
  • Controller comparison
  • Pumping Speed Curve
  • Controller outline

Specifications

Pumping speed (N2)75 I/s (V80) / 77 I/s (81-T)
Pumping speed (He)55 I/s (V80) / 65 I/s (81-T)
Ultimate pressure (with rotary)< 5x10e-9 mbar
Rotational speed60 KRPM (V80) / 80 KRPM (81-T)
Power consumption90 W (max 500)
Noise level< 50 dB (A)
Weight (ISO 63)1.6 Kg

Frequently asked questions

What are the main advantages of the Turbo 81-T pump?

The Turbo 81-T pump offers better performance in a smaller package with optimized fins for lower operating temperatures.

Can the existing V80 controller be replaced?

Yes, the existing non-compatible controller can be replaced with the 81-AG series controller.

Which accessories need to be replaced during the upgrade?

All accessories except the inlet screen must be replaced.