Agilent 500 Ion Trap LC/MS Technical Overview

The document describes the features and characteristics of the Agilent 500 Ion Trap LC/MS, including the Enhanced Charge Capacity (ECC)


Agilent 500 Ion Trap LC/MS Technical Overview - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
6
File format
PDF
File size
3 MB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
D53B892970FB33C90BB85B95D274AF78

About this manual

This document provides a technical overview of the Agilent 500 Ion Trap LC/MS with Enhanced Charge Capacity (ECC). It explains how modifying the electrical multipole content of the trapping field allows the instrument to trap a significantly higher number of ions compared to competitive models. This enhancement is achieved through additional mechanically formed octapole and electrically generated hexapole components in the trapping field.

What's inside

  • Introduction
  • Experimental Conditions
  • Discussion and Results
  • Benefits
  • References

Frequently asked questions

What is the primary benefit of the Enhanced Charge Capacity (ECC) feature in the Agilent 500 Ion Trap LC/MS?

The Enhanced Charge Capacity allows the instrument to reliably detect trace amounts of analytes even when they are present in a high concentration of coeluting analytes within heavy matrices.

How much can the ion time be increased on the Agilent 500 Ion Trap before mass resolution is noticeably degraded?

The Agilent 500 Ion Trap can tolerate a 100-fold increase in the ion time with little noticeable decrease in mass resolution.

What is the maximum mass-to-charge ratio of the ions used in the mass calibration compound Ultramark?

The mass calibration compound Ultramark contains ions with mass-to-charge ratios up to 2200.