Agilent Polynuclear Aromatic Hydrocarbons by HPLC Application Note

This application note describes reproducible methods for sample preparation and HPLC analysis of PNAs in soil, barbecued meat and waste water, all with good recovery rates, and detection limits that exceed regulatory requirements.


Agilent Polynuclear Aromatic Hydrocarbons by HPLC Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
258 KB
Published
03 July, 2012
Updated
30 September, 2023
Model
HP 1050
MD5 checksum
1669CC8EEABC38A315B235576D0BE1D4

About this manual

This document details methods for the analysis of polynuclear aromatic hydrocarbons (PNAs) using High-Performance Liquid Chromatography (HPLC). It covers sample preparation techniques for various matrices, including soil, barbecued meat, and waste water, aiming for good recovery rates and detection limits that meet regulatory requirements.

The note evaluates two specific HPLC configurations: a standard-bore system using the HP 1050 Series with a 4.6-mm id column, and a narrow-bore system using the HP 1090 Win Series with a 2.1-mm id column. Both setups utilize fluorescence and UV-Visible diode-array detectors to optimize sensitivity and selectivity for different compounds.

What's inside

  • Introduction
  • Sample Preparation
  • Chromatographic Method
  • Separation of PNAs
  • Optimization of Fluorescence Detection
  • Optimization of UV-Visible Absorbance
  • Statistical Analysis
  • Results and Discussion

Specifications

Detection limit (Fluorescence)0.3 picograms
Detection limit (UV-Visible)50 pg
Column ID (Standard-bore)4.6 mm
Column ID (Narrow-bore)2.1 mm
Flow rate (Standard-bore)1.5 ml/min
Flow rate (Narrow-bore)0.42 ml/min
Column temperature28°C

Frequently asked questions

Which Agilent software is used for automation?

The entire HPLC method, including separation and report generation, can be automated using the HPLC ChemStation (DOS Series).

What is the minimum detection limit for acenaphthylene using UV-Visible absorbance?

The method can identify acenaphthylene down to 44 pg.

What is the flow rate used for the narrow-bore column system?

The narrow-bore column system operates at a flow rate of 0.42 ml/min.

What temperature is recommended for the column thermostat to ensure reproducible separations?

A column thermostatting temperature of 28°C ensures the most reproducible separations.