Agilent 6890/5973 GC/MSD System Application Note

The analysis of pesticides in water, foods, and soil is important worldwide. This document discusses the changes and regulations in environmental monitoring in the United States, Europe, Asia-Pacific region, and Latin America. It highlights the challenges and limitations of using gas chromatography/mass spectrometry (GC/MS) for pesticide detection, and the reliance on gas chromatography with an electron capture detector (GC/ECD). It also introduces the new 6890/5973 GC/MS system, which offers lower detection limits and can meet the current demands of environmental monitoring.


Agilent 6890/5973 GC/MSD System Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
432 KB
Published
02 July, 2012
Updated
29 September, 2023
MD5 checksum
836FA95E2890492CF4F787015389152E

About this manual

The document details the analysis of pesticides in water, foods, and soil using the Agilent 6890/5973 GC/MSD system. It highlights the system's ability to achieve detection limits well below 10 picograms on-column, surpassing previous GC/MS capabilities and offering an alternative to GC/ECD systems. The analysis utilized Electron Ionization (EI) with both Selected Ion Monitoring (SIM) and scan data collection modes.

What's inside

  • Abstract
  • Introduction
  • Optimized Instrumentation
  • 6890 GC
  • 5973 MSD
  • MS Productivity ChemStation
  • Results
  • Lowered Detection Limits

Specifications

Scan sensitivity specification1 picogram of octafluoronaphthalene
HED detector voltage10 kV
Detection limit (SIM mode)Well below 10 pg on-column

Frequently asked questions

What detection limits does the 6890/5973 GC/MSD achieve for pesticides?

The repeatability-derived detection limit in SIM mode well exceeds 10-picogram (on-column) detectability for a group of fifteen pesticides.

How does the HED detector improve sensitivity?

The HED detector improves sensitivity by increasing the efficiency of ion detection, particularly for ions with m/z > 200.