Agilent Prep-C18 Preparative Scaleup of Calcium Channel Blockers Application Note

Agilent Reversed Phase Prep columns demonstrated both good loadability and good scalability for calcium channel blockers using gradient conditions. The columns also dis- played good resolution at all dimensions and allowed a linear scaleup from analytical to preparative dimensions


Agilent Prep-C18 Preparative Scaleup of Calcium Channel Blockers Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
459 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
201031E8905227E01666924FA533E2F1

About this manual

This document describes a study on the preparative scaleup of calcium channel blockers (Nifedipine, Nimodipine, and Nisoldipine) using Agilent Reversed Phase Prep columns. The authors demonstrate that the columns exhibit good loadability and scalability, allowing for a linear scaleup from analytical to preparative dimensions while maintaining resolution.

The study focuses on the Agilent Prep 5-um Scalar column and the Agilent Prep-C18 series. It outlines the theoretical considerations for scaleup, including the relationship between sample capacity and column dimensions, and presents data on column loadability and scalability across various column internal diameters (4.6 mm to 50.8 mm).

What's inside

  • Abstract
  • Introduction
  • Pharmaceuticals
  • Analytical Separation
  • Theoretical Considerations
  • Column Loadability
  • Scalability
  • Chromatographic Conditions

Specifications

Column TypeAgilent Prep-C18
Particle Size5-um
Column Length100 mm
Detector Signal345 nm
Sample Concentration80 mg/mL

Frequently asked questions

What are the three calcium channel blockers separated in this study?

The study separates Nifedipine, Nimodipine, and Nisoldipine.

What is the recommended resolution threshold for stopping the scaleup process?

The scaleup process should stop when resolution drops below 1.2.

What solvent is used for the sample stock solution?

The sample stock solution is prepared in dimethylsulfoxide (DMSO).