Agilent 1100 Series Mass-based Fraction Collection Application Note

This Application Note demonstrates the mass-based purification of low gram amounts of compounds using the Agilent 1100 Series purification platform. Starting the experiments on 21.2-m i.d. columns two application examples were scaled up to a 50-mm i.d. column operated at 100 mL/min. Fractions with purities of 85 and 95 % were obtained by mass-based fraction collection even though the column overload did not allow the UV-trace to show separated peaks.


Agilent 1100 Series Mass-based Fraction Collection Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
397 KB
Published
02 July, 2012
Updated
01 October, 2023
MD5 checksum
B110D57A4A5E48FC2AD261A13DCC7FB5

About this manual

The document describes a method to purify up to 4 grams of crude product in a single run on 50 mm diameter columns operated at 100 mL/min. It utilizes mass-based fraction collection mode, which allows for the isolation of compounds with high purity even when UV-trace peaks are not separated due to column overload.

The text details the specific equipment configuration required, including preparative pumps, a dual-loop autosampler, and a multiwavelength detector. Two test mixtures containing calcium channel blockers were used to validate the method on both analytical-scale (21.2 mm i.d.) and preparative-scale (50 mm i.d.) columns.

What's inside

  • Abstract
  • Introduction
  • Equipment
  • Results and discussion
  • System configuration
  • Purification of gram amounts
  • Conclusion

Frequently asked questions

What is the maximum amount of crude product that can be purified in a single run on a 50 mm diameter column?

The system can purify up to 4 grams of a crude product in a single run on 50 mm diameter columns.

What flow rate was used for the high flow rate experiments on 50 mm diameter columns?

The experiments were operated at a flow rate of 100 mL/min.

What purity was achieved for nimodipin when purified from test mix A using the large-scale column?

The re-analysis of the collected fraction presented a purity of 96.2% for nimodipin.