Agilent 2100 Bioanalyzer Stringent RNA Quality Control Application Note

This application note describes the use of the Agilent 2100 bioanalyzer for stringent RNA quality control and the advantages of this method for high-resolution analysis of small and very small RNA samples.


Agilent 2100 Bioanalyzer Stringent RNA Quality Control Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
296 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
FC30EBF4E9A193DA82674EC8A7F54867

About this manual

The document explains that successful RNA technology relies on high-quality RNA samples. It introduces the Agilent 2100 bioanalyzer as a chip-based system using pre-packed kits to reduce manual intervention and improve data accuracy compared to traditional gel electrophoresis.

A key focus of this note is the importance of heating RNA samples prior to electrophoresis. The text details how heating destabilizes base-pairing in fragmented rRNAs and aggregates, preventing misleading results where broken fragments mimic intact molecules. It provides examples from human, fish, and plant tissues.

What's inside

  • Introduction
  • Materials and methods
  • Results
  • Heating the RNA sample
  • Table 1: Length variations of rRNAs

Frequently asked questions

How much total RNA is required for analysis using the RNA 6000 Nano LabChip kit?

Only about 25 ng of total RNA are required for analysis.

What are the specific incubation conditions for heating RNA samples prior to electrophoresis?

RNA samples should be incubated for 2 minutes at 70 °C in thin-wall 0.5 mL PCR tubes.

What is the expected migration time difference between 18S and 28S rRNAs for heated human RNA samples?

The migration difference is about 5.6 seconds for heated human RNA samples.