Agilent 5100 Protein 200 HT-2 Assay Antibody Analysis Application Note

This Application Note describes the analysis of different antibody sam- ples under reducing and non-reducing conditions.


Agilent 5100 Protein 200 HT-2 Assay Antibody Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
3.7 MB
Published
02 July, 2012
Updated
30 September, 2023
Model
HT-2
MD5 checksum
97E1657832DDF97E0E0FBF4A6E2822D7

About this manual

This document describes a suitability test for the Agilent 5100 Automated Lab-on-a-Chip Platform (ALP) using the Protein 200 HT-2 Assay. The study focuses on analyzing different antibody samples to determine linearity, dynamic range, sizing reproducibility, and quantification limits. The analysis is performed under both reducing and non-reducing conditions to assess the system's performance compared to traditional methods like SDS-PAGE and SEC.

What's inside

  • Abstract
  • Introduction
  • Material and methods
  • Results and discussion
  • Linearity of antibody quantitation
  • Sizing reproducibility
  • Limit of quantification (LOQ)
  • System robustness

Specifications

Size range15 to 220 kDa
Chip capacityup to 2000 samples
Chip part numberG3011-87300
Reagent kit part numberG3011-68750
Assay specification CV3 % CV

Frequently asked questions

What is the linear dynamic range for non-reduced antibody 1?

A good linearity (R² ≥ 0.98) is achieved up to 1000 pg/mL.

What is the linear dynamic range for reduced antibody 2?

A good linear dynamic range is achieved up to 2000 pg/mL.

What is the limit of quantification (LOQ) for antibody 1 in non-reduced form?

The LOQ is 10 pg/mL, determined by a signal to noise greater than 6.

What is the limit of quantification (LOQ) for antibody 2 in reduced form?

The LOQ is 20 pg/mL, determined by a signal to noise greater than 6.