Agilent More speed better resolution lower LOD using liquid chromatography fluorescence detection - Comparing the 1100 Series LC to the 1200 Series Rapid Resolution system

This document compares the performance differences in analyzing vitamins B2 and B6 using liquid chromatography and fluorescence detection. It compares the Agilent 1100 Series LC system with the Agilent 1200 Series Rapid Resolution LC system. The new column technology significantly improves performance, including faster run time, lower detection limit, and better resolution.


Agilent More speed better resolution lower LOD using liquid chromatography fluorescence detection - Comparing the 1100 Series LC to the 1200 Series Rapid Resolution system - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
169 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
10A97CE4872EB65DFC4C32B7B6412E2A

About this manual

This document compares the performance of the Agilent 1100 Series LC system with the newer Agilent 1200 Series Rapid Resolution LC (RRLC) system. The comparison focuses on the analysis of vitamins B2 and B6 using fluorescence detection, demonstrating improvements in speed, resolution, and limit of detection (LOD) achieved with the new system and sub-2-micron particle columns.

The text details the specific equipment used, including the Agilent 1200 Series binary pump, autosampler, thermostat, and fluorescence detector. It presents chromatographic conditions and results showing that the 1200 Series system reduces run time from 10 minutes to approximately 3 minutes while significantly lowering the LOD for vitamin B2 from 20 pg to 1.1 pg.

What's inside

  • Abstract
  • Introduction
  • Equipment
  • Chromatographic conditions
  • Results and discussion
  • Conclusion
  • References

Specifications

Column particle size1.8 pm
Column dimensions50 x 2.1 mm
Column temperature30°C
Flow rate0.7 mL/min
Run time3 min
LOD for Vitamin B21.1 pg
LOD for Vitamin B6< 25 pg
Resolution for Vitamin B6> 5

Frequently asked questions

What is the run time for the new Agilent 1200 Series system?

The run time is only about 3 minutes.

What is the run time for the vitamin analysis using the Agilent 1200 Series RRLC system?

The run time is only about 3 minutes when using the new column technology.

What is the limit of detection for vitamin B2 using the Agilent 1200 Series RRLC system?

The limit of detection for vitamin B2 is 1.1 pg with a signal-to-noise ratio of 2.

What is the resolution for the B6 vitamins when using the Agilent 1200 Series RRLC system?

The resolution for vitamin B6 is greater than 5.