Agilent 1100 Series LC/MSD Trap Ion Trap Quantitative Analyses Application Note

This note describes the effect of coeluting internal standard concentration on the dynamic range of Ion Trap quantitative analyses and demonstrates the advantages of Ion Trap technology over triple-quadrupole instruments for quantitative applications in high-throughput environments.


Agilent 1100 Series LC/MSD Trap Ion Trap Quantitative Analyses Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
5
File format
PDF
File size
65 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
42EADA16D3D999E178C82BFAFE306C0D

About this manual

This document is an application note by Agilent Technologies demonstrating the performance of the Agilent 1100 Series LC/MSD Trap ion trap mass spectrometer. It specifically addresses the effect of coeluting internal standards on quantitative analyses, showing that modern ion trap designs maintain linearity over three orders of magnitude even with vastly excessive concentrations of internal standards.

What's inside

  • Introduction
  • Experimental
  • Analysis Method
  • Results and Discussion
  • Conclusions
  • Authors

Specifications

Mobile phase A1% Acetic acid (Aq)
GradientIsocratic at 15% B

Frequently asked questions

What is the dynamic range of the LC/MSD Trap for quantitative assays?

The LC/MSD Trap demonstrates linearity over three orders of magnitude for quantitative assays.

Does the presence of coeluting internal standards affect instrument performance?

No, the presence of increasing amounts of coeluting internal standard has no effect on analytical linearity or instrument performance.

What are the advantages of full scan MS/MS mode?

Full scan MS/MS mode provides structural information for reliable compound identification and allows quantitation ions to be selected post-acquisition.