Agilent 1100 Series LC/MSD Trap Ion Trap Mass Spectrometer Application Note

This application note describes the use of an Ion Trap LC/MS/MS intelligent database search to identify phosphorylation sites in peptides resulting from proteolytic digestion of a phosphorylated protein


Agilent 1100 Series LC/MSD Trap Ion Trap Mass Spectrometer Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
286 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
1B05B62C7FE9A599B4737EE48F8D44BA

About this manual

This document describes an application using automated tandem mass spectrometry coupled to online capillary liquid chromatography to identify post-translational modifications, specifically phosphorylation sites in proteins. The technique utilizes enzymatic digestion, capillary LC separation, and data-dependent scan functions to select parent ions for collision-induced dissociation (CID).

What's inside

  • Introduction
  • Experimental
  • Results and Discussion
  • Analysis Method
  • MS Conditions
  • Database Search
  • Conclusions

Specifications

ColumnZorbax 300SB-C18, 150 x 0.3 mm, 5 µm
Flow rate5.5 µl/min
Mobile phase A0.1% formic acid in water
Mobile phase B0.1% formic acid in acetonitrile
Ionization modePositive ESI
Nebulizer pressure15 psig
Automatic MS/MSNumber of parents: 2
Isolation width3 m/z
Fragmentation amplitude1.0 V

Frequently asked questions

What software was used for database searching?

The Mascot database search program was used to search the uninterpreted MS/MS spectra.

Which enzyme was used for protein digestion?

Beta-casein was digested with trypsin in ammonium bicarbonate buffer.

What search parameters were set for phosphorylation?

Search parameters were set to consider possible phosphorylation at all serine, threonine, or tyrosine residues.

Download PDF (286 KB)