Agilent High-Throughput Protein Quantitation Using MRM Application Note

Quantitative proteomics using multiple reaction monitoring (MRM) has emerged as an important methodology for biomarker validation. Multiple-reaction monitoring on a triple quadruple (QQQ) mass spectrometer provides superior sensitivity and selec- tivity for targeted peptides in a complex sample. MRM also offers high precision in quantitation and a fast scan speed, which makes it an ideal technology for validating biomarkers in a high-throughput fashion. This application note demonstrates the entire workflow for biomarker validation offered by Agilent Technologies using multiple reaction monitoring together with statistical analysis software. Thousands of peptide transitions were generated using Agilent Q-TOF and Spectrum Mill MRM selector. The transition list was imported into MassHunter acquisition software equipped with dynamic MRM functionality and monitored in a single LC/MS run. Finally, MRM quantitation results were analyzed using Mass Profiler Pro for principle component analysis (PCA), hierarchical clustering and ANOVA analysis.


Agilent High-Throughput Protein Quantitation Using MRM Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
712 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
5436A6034B0A8D658B596106CC57C94F

About this manual

This document describes a workflow for quantitative proteomics and biomarker validation using Multiple Reaction Monitoring (MRM) on Agilent mass spectrometers. The study utilizes a triple quadrupole (Q0Q) mass spectrometer interfaced with an HPLC-Chip for nanoflow LC/MS analysis to achieve high sensitivity and selectivity.

The workflow involves sample preparation of human plasma, depletion of abundant proteins, and digestion. Data acquisition is performed using dynamic MRM functionality in MassHunter software, allowing thousands of peptide transitions to be monitored in a single run. Results are analyzed using Mass Profiler Pro for statistical analysis, including principal component analysis and hierarchical clustering.

What's inside

  • Abstract
  • Introduction
  • Sample Preparation
  • LC/MS Analysis
  • Software
  • Results and Discussion
  • MRM Selector Parameters
  • Reproducibility

Specifications

Analytical column dimensions150 x 0.075 mm
Enrichment column volume40 nL
Injection volume1 pL
Analytical pump flow rate300 nL/min
Loading pump flow rate3 L/min
Drying gas flow5 L/min
Capillary voltage1,800 V
Collision energy slope3.6
MS response RSDbelow 5%
Retention time RSDless than 0.04 min

Frequently asked questions

What software is used for dynamic MRM acquisition?

MassHunter acquisition software equipped with dynamic MRM functionality is used to monitor the transition list.

Which software is used for statistical analysis of MRM results?

Mass Profiler Pro is used to perform principal component analysis, hierarchical clustering, and ANOVA analysis.

How many data points are recommended across peaks in dynamic MRM?

The cycle time is set to obtain at least 15 data points across the peaks.