Agilent Metabolic Stability Study Using Cassette Analysis and Polarity Switching in a UHPLC-Triple Quad LC/MS System App

This application note demonstrates the performance of the Agilent 1290 Infinity LC System coupled to an Agilent 6460A Triple Quadrupole LC/MS for the evaluation of metabolic stability using the cassette analysis approach combined with fast polarity switching.


Agilent Metabolic Stability Study Using Cassette Analysis and Polarity Switching in a UHPLC-Triple Quad LC/MS System App - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
236 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
CC0FB8E83454A2211B9E4A5D19E59264

About this manual

The document presents a UHPLC/MS/MS method that achieves high throughput analysis with retention times less than 0.8 minutes at a flow rate of 1.5 mL/min and binary pump pressure of 1,100 bar. It utilizes the Agilent 1290 Infinity LC System and an Agilent 6460A Triple Quadrupole LC/MS System with Jet Stream technology to monitor six transitions while switching polarity in just 30 ms.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Reagents and supplies
  • Incubation of substrates
  • Equipment
  • Agilent 1290 Infinity Methods
  • Results and Discussion
  • Speed and data quality

Specifications

Flow rate1.5 mL/min
Binary pump pressure1,100 bar
MS cycle time114 ms
Polarity switching time30 ms
Data points per peakat least 9 points
Accuracy range90.7 — 107.1%

Frequently asked questions

What is the polarity switching time for this method?

The polarity switching time is 30 ms.

How many data points are collected across chromatographic peaks?

At least 9 points are collected across the chromatographic peaks.

What is the accuracy range for concentration measurements?

The accuracy over the expected concentration range is between 90.7% and 107.1%.