Agilent 7700x ICP-MS Application Note for Drinking Water Analysis

This document introduces the advantages of Agilent 7700x ICP-MS technology in drinking water analysis. This technology uses the 3rd generation Octopole Reaction System (ORS3) technology to effectively remove polyatomic interferences. He mode is simple and easy to use, and the 7700x ICP-MS can remove all polyatomic interferences in a single set of conditions without requiring any prior knowledge about the sample, and without the need for complex mathematical corrections. This makes setup easier, detection limits are lower, analysis is faster, and interference removal is improved, especially for the difficult plasma-based interferences affecting iron and selenium. Currently, Se detection limits of less than 20ppt are routinely achievable in He mode.


Agilent 7700x ICP-MS Application Note for Drinking Water Analysis - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
538 KB
Published
02 July, 2012
Updated
30 September, 2023
Model
7700x
MD5 checksum
EF7A3088E29663432D899EC4C4A6A960

About this manual

The Agilent 7700x ICP-MS utilizes 3rd generation Octopole Reaction System (ORS) technology to improve the performance of helium collision mode for removing polyatomic interferences in environmental analysis. This system allows for the removal of all polyatomic interferences under a single set of conditions without requiring complex mathematical corrections or previous knowledge of the sample.

The document presents data demonstrating the instrument's sensitivity and accuracy, showing method detection limits for various elements in drinking water. It highlights the ability to achieve selenium detection limits less than 20 parts per trillion (ppt) and discusses the robustness of the High Matrix Introduction (HMI) system for analyzing samples with high total dissolved solids.

What's inside

  • Introduction
  • Better Sensitivity, Better Accuracy, Better Dynamic Range
  • Typical method detection limits
  • CCV recoveries
  • NIST 1643e reference water analysis
  • Conclusions

Specifications

Selenium detection limit (He mode)less than 20 ppt
Vanadium BEC (He mode)7.7 ppt

Frequently asked questions

Does the 7700x ICP-MS require reactive cell gases for drinking water analysis?

No, the 7700x ICP-MS removes all polyatomic interferences under a single set of conditions without requiring reactive cell gases.

What is the operating dynamic range of the 7700x ICP-MS?

It has the largest operating dynamic range of any quadrupole ICP-MS.