Agilent 1100 Series HPLC System Application Note for PNA Analysis

This is an application note about improving data quality in the automated HPLC analysis of PNAs (PAHs)


Agilent 1100 Series HPLC System Application Note for PNA Analysis - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
767 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
27C313D06A5F1E2D65B05FCB2C811ACD

About this manual

This document presents an optimized method for the automated HPLC analysis of polycyclic aromatic hydrocarbons (PNAs), also known as PAHs. The study focuses on improving data quality by addressing issues with retention time stability and fluorescence detection sensitivity. Key improvements include the use of vacuum degassing to prevent quenching effects and thermostatting the column compartment slightly below ambient temperature to enhance resolution of critical peak pairs.

The authors utilized an Agilent 1100 Series HPLC system configured with a quaternary gradient pump, vacuum degasser, autosampler, thermostatted column compartment, diode-array detector, and fluorescence detector. The method achieved retention-time stability within 0.05–0.2% RSD and peak area precision below 2% in the low ng range. Additionally, the improved diode-array detection system allowed for the detection of PNAs at ppb levels with automatic identification via UV-Visible spectral library search.

What's inside

  • Introduction
  • Experimental
  • Impact of column temperature on separation
  • Precision of retention times and peak areas
  • Impact of vacuum degassing on fluorescence detection
  • Spectral information with diode-array detection
  • Automated positive identification of PNAs

Specifications

Column internal diameter2.1 mm
Solvent flow rate0.4 ml/min
Retention time stability0.05 — 0.2 % RSD
Peak area precision (low ng range)below 2 %
Detection limitppb levels
Identification thresholdless than 1 mAUFS

Frequently asked questions

How does vacuum degassing improve PNA analysis?

Thorough degassing prevents quenching effects in fluorescence detection and improves retention-time stability.

What is the effect of column temperature on PNA separation?

Thermostatting the column temperature slightly below ambient temperature improves the resolution of critical peak pairs.

Can PNAs be identified automatically using this method?

Yes, an improved diode-array detection system allows identity confirmation automatically by UV-Visible spectral library search.