Agilent 7000 Series Triple Quadrupole Mass Spectrometer Pressure Controlled Tee Application Note

This note describes the Pressure Controlled Tee (PCT) configuration for rapid and universal GC/MS backflushing applied to selected reaction monitoring experiments on the Agilent 7890 GC used with the Agilent 7000 Series Triple Quadrupole Mass Spectrometer. This means that rapid backflushing and quick column and inlet maintenance without venting the MS system can be incorporated without a loss in detection limits using a simple PCT configuration.


Agilent 7000 Series Triple Quadrupole Mass Spectrometer Pressure Controlled Tee Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
1.1 MB
Published
02 July, 2012
Updated
01 October, 2023
MD5 checksum
B641C57A94ABA4D0C42E673AAC4CB86E

About this manual

This document presents the results of an experiment using the Pressure Controlled Tee (PCT) configuration for rapid and universal GC/MS backflushing on the Agilent 7890 GC coupled with the Agilent 7000 Series Triple Quadrupole Mass Spectrometer. The study aimed to determine if adding this plumbing configuration would compromise signal sensitivity or detection limits in multiple reaction monitoring (MRM) mode.

The results indicate that implementing the PCT configuration maintains signal height and area without degrading compound retention time reproducibility or other performance metrics. Consequently, rapid backflushing and service of the injection port and column can be introduced without compromising analytical integrity or requiring the mass spectrometer to be vented.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Results
  • Conclusions
  • Injection Port Diagram
  • Comparison Table
  • References

Specifications

Target Compoundoctafluoronaphthalene (OFN)
Concentration100 fg
Column 2 Length15 m
Carrier Gashelium
Post run Temperature200 °C

Frequently asked questions

Does using a Pressure Controlled Tee affect sensitivity?

No, the results show no loss in sensitivity as measured for peak area, height, or signal-to-noise ratio.

Does the PCT configuration require venting the MS system?

No, rapid backflushing and service can be performed without venting the MS system.

What is the impact on retention time reproducibility?

Retention time reproducibility is unaffected, achieving a standard deviation of less than 0.1 seconds.