Agilent 1100 Series Analysis of Bendiocarb Metabolite by HPLC

This document describes the HPLC analysis method of bendiocarb and metabolite


Agilent 1100 Series Analysis of Bendiocarb Metabolite by HPLC - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
151 KB
Published
02 July, 2012
Updated
01 October, 2023
MD5 checksum
AADF45B978053626D5DD0E51EC6A4EC6

About this manual

This document details a High-Performance Liquid Chromatography (HPLC) method for the analysis of the insecticide bendiocarb and its metabolite. The method utilizes a Hypersil ODS C18 column with a mobile phase of water and acetonitrile, detected at 212 nm using a diode array detector. The procedure covers extraction from soil and water, as well as separation and detection parameters.

What's inside

  • Abstract
  • Separation
  • UV-visible detection
  • Conditions
  • Sample preparation
  • Equipment

Specifications

Column100 x 2.1-mm Hypersil ODS C18, 5 ym
Mobile phaseWater—acetonitrile (65:35 isocratic mixture)
Flow rate0.36 ml/min
Temperature40 °C
Detection wavelength212 nm (16 nm bandwidth)
Detection limit4 yg/!

Frequently asked questions

What equipment is used for sample preparation?

Samples can be extracted from soil using Soxhlet equipment or by ultrasonic treatment in solution, and from water using liquid—solid or liquid-liquid techniques.

What is the detection limit for this method?

The detection limit is 4 yg/! without sample enrichment.