Agilent 5890 Series II Optimized Analysis of Gasoline BTEX in Water Soil Using GC/FID with Purge Trap Application Note

Optimized analysis of gasoline (BTEX) in water and soil using GC/FID with purge and trap


Agilent 5890 Series II Optimized Analysis of Gasoline BTEX in Water Soil Using GC/FID with Purge Trap Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
144 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
893BF52CC5CDBDD620E303F978BFE236

About this manual

This document describes an optimized method for determining gasoline components, specifically gasoline range organics (GROs) and benzene, toluene, ethylbenzene, and xylenes (BTEX), in contaminated water and soil. The method utilizes modified EPA Methods 8015/8020 with capillary gas chromatography (GC) and a flame ionization detector (FID) or photoionization detector (PID). The analysis employs an Agilent 7695A Purge and Trap (P and T) system coupled with an Agilent 5890 Series II gas chromatograph.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Results and Discussion
  • Optimized GC Run Time
  • Optimized P and T Cycle Time
  • Sample Purge
  • Dry Purge
  • Desorption
  • Table 1. Instrument Requirements and Optimized Conditions

Specifications

Gas chromatograph5890 Series II
ColumnHP-1, 30 m x 0.53 mm x 5.0 µm
DetectorFID
Purge time11 min
Dry purge time1 min
Desorb time2 min
Bake time5 min

Frequently asked questions

What is the optimal sample purge time for gasoline and GROs analysis?

The optimal sample purge time is 11 minutes, or 440 ml of helium purge gas, to ensure efficient recovery of high-boiling fractions.

What is the optimal GC run time achieved in this study?

The optimized GC run time achieved is 17 minutes using a thin-film HP-1 column and faster oven temperature programs.