Agilent Application Note: Determination of Pharmaceuticals in Water by SPE and LC/MS/MS

This document describes the analysis of 19 pharmaceuticals in positive ion mode and 11 pharmaceuticals in negative ion mode using solid phase extraction (SPE) and liquid chromatography/tandem mass spectrometry (LC/MS/MS). The analysis was performed at low picogram level on column without any derivatization. Good linearity was observed for analytes from 1 pg to 1 ng on column. Repeatability from six injections of analytes at 5 pg on column showed RSDs below 15%, for all target compounds except for fluoxetine at 23%.


Agilent Application Note: Determination of Pharmaceuticals in Water by SPE and LC/MS/MS - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
117
File format
PDF
File size
34.7 MB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
37C031883A319206D1E947EEDD8728BC

About this manual

The document details an analytical method developed by the U.S. Geological Survey National Water Quality Laboratory, implemented with Agilent Technologies instrumentation to detect trace pharmaceuticals in surface and ground water. The study focuses on monitoring man-made chemicals that disrupt endocrine systems.

Specific procedures for sample preparation using Oasis HLB solid-phase extraction (SPE) are provided. The analysis utilizes a ZORBAX Extend-C-18 column coupled with an Agilent 1200 LC system and G6410A Mass Spectrometer, operating in both positive and negative ion modes via Electrospray Ionization (ESI).

What's inside

  • Introduction
  • Instrumentation
  • Sample Preparation Procedure
  • Positive Ion Mode MRM Parameters
  • Negative Ion Mode MRM Parameters
  • Results and Discussion

Frequently asked questions

How many pharmaceuticals were analyzed in positive ion mode?

The method analyzes 19 pharmaceuticals using the positive ion mode.

What is the lowest concentration level at which analytes were analyzed on column?

Analytes were successfully analyzed at a low picogram level of 1 pg on column.

What was the repeatability result for fluoxetine?

The repeatability for fluoxetine showed an RSD of 23%.