Agilent 4500 ICP-MS Technical Note: Determination of As in High Chloride Samples

This document introduces the method for determining As in samples containing high concentration of chloride using Agilent 4500 ICP-MS.


Agilent 4500 ICP-MS Technical Note: Determination of As in High Chloride Samples - cover page
Brand
Agilent
Category
Other
Document type
Other
Language
English
Pages
4
File format
PDF
File size
80 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
636F37CE7C0CF8CC62C22E20CA41DE63

About this manual

This document addresses the spectral interference of ArCl on Arsenic (As) determination in samples containing high levels of chloride. It explains that while high-resolution mass spectrometry can separate these ions, it compromises detection limits. The 4500 ICP-MS overcomes this by using mathematical correction via elemental equations resident in the ChemStation software.

What's inside

  • Introduction
  • Elemental equations
  • Influence of Cl Concentration
  • Determination of As in Cl Matrix
  • Operating conditions

Specifications

Resolving power required for separation>7500
Ion signal stability<1%RSD over 2 hours
Spray chamber temperature control+/-0.1 °C
RF power13kW
Sampling depth8 mm
Plasma gas flow16 l/min
Auxiliary gas flow1.0 Vmin
Carrier gas flow21.15 Vmin
Nebulizer typeConcentric type

Frequently asked questions

How does the 4500 ICP-MS correct for ArCl interference on Arsenic?

It uses elemental equations resident in the ChemStation software to correct for interferences on both As and Se simultaneously.

Can the 4500 ICP-MS determine Arsenic in 5% HCl?

Yes, it can determine As at the ppb and sub-ppb level even in 5% HCl using interference correction.

What is the concentration range for As determination in 5% HCl?

The document demonstrates determination from 0 to 200 µg/l.