Agilent 6520 Accurate-Mass Q-TOF LC/MS Peptide Mapping Application Note

Peptide Mapping of a Monoclonal Antibody using a Microfluidic-based HPLC-Chip coupled to an Agilent Accurate-Mass Q-TOF LC/MS


Agilent 6520 Accurate-Mass Q-TOF LC/MS Peptide Mapping Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
513 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
4BF5E34EE2E54A8877AC7139EC09EC01

About this manual

The document details a method for sequence characterization of monoclonal antibodies (mAbs) at low nanogram levels. It utilizes an Agilent HPLC-Chip system coupled with an Agilent Accurate-Mass Quadrupole Time-of-Flight (Q-TOF) LC/MS instrument. The high performance and mass accuracy of the platform, combined with MassHunter and BioConfirm software, enable complete sequence coverage and mass confirmation of peptides.

What's inside

  • Abstract
  • Introduction
  • Materials
  • Sample pretreatment
  • Protease digestion
  • Instrumentation
  • LC Parameters
  • MS Parameters
  • Data analysis
  • Results and discussion

Specifications

HPLC-Chip Column5 ym, ZORBAX 300SB-C18 (300A), 40 nL enrichment column and a 75 mm x 43 mm analytical column
Flow rate (enrichment)3 L/min
Flow rate (analytical)600 nL/min
Sample injected50 ng
Mass accuracy4 ppm
Sequence coverage (trypsin)95% heavy chains, 85% light chains

Frequently asked questions

What software is used for data analysis?

Data was processed using Agilent MassHunter Qualitative Analysis software, Agilent MassHunter BioConfirm and Agilent Spectrum Mill MS Proteomics Workbench software packages.

What is the flow rate for the enrichment column?

The flow rate from the Agilent 1200 Series Capillary Pump to the enrichment column is 3 L/min.