Agilent 6460 Triple Quadrupole LC/MS Determination of Acidic Herbicides Application Note

This application note describes the use of an Agilent 6460 triple quadrupole LC/MS equipped with Agilent Jet Stream technology and direct aqueous injection for the analysis of a suite of acidic herbicides in several water matrices, thereby negating the need for solid phase extraction. Standard deviations in the range of 2.5% to 6.6% for analyte precision and recoveries of between 91.5% and 105.1% were observed over 11 batches of samples. Limits of detection were less than 10 ng/L (10 ppt) for all the compounds in the suite thus meeting current industry standards.


Agilent 6460 Triple Quadrupole LC/MS Determination of Acidic Herbicides Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
2.5 MB
Published
02 July, 2012
Updated
01 October, 2023
MD5 checksum
A1B8979B8296EA0D07C4217D081CBF82

About this manual

The document presents an analytical method for detecting a suite of acidic herbicides in potable and environmental water samples. The method utilizes direct aqueous injection, eliminating the need for time-consuming solid phase extraction (SPE) steps typically required for these compounds.

Validation was performed on 11 batches of samples including borehole groundwater, potable water, river water, and deionized water. The results demonstrated recoveries between 91.5% and 105.1%, standard deviations ranging from 2.5% to 6.6%, and limits of detection less than 10 ng/L (10 ppt) for all compounds in the suite.

What's inside

  • Introduction
  • Experimental
  • Sample Preparation
  • Instrumentation
  • LC Conditions
  • Triple Quadrupole MS Conditions
  • MRM Parameters
  • Results and Discussion
  • Validation
  • Conclusions

Specifications

Column dimensions2.1 mm x 100 mm
Column temperature60 °C
Total run time26.0 min
Gas temperature (Negative)250 °C
Veap voltage (Negative)2000 V

Frequently asked questions

Does this method require solid phase extraction?

No, the analysis is performed by direct aqueous injection, negating the need for solid phase extraction.

What are the limits of detection for the compounds?

Limits of detection were less than 10 ng/L (10 ppt) for all the compounds in the suite.

What is the range of analyte recoveries observed?

Recoveries ranged between 91.5% and 105.1% over 11 batches of samples.