Agilent GC-ICP-MS Application Note for Sulfur Quantification in Gasoline

This application note examines the quantification of total sulfur levels (using compound independent calibration) and sulfur speciation in reformulated gasoline.


Agilent GC-ICP-MS Application Note for Sulfur Quantification in Gasoline - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
561 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
8A4C0E2C34C50D0FE8D396AF4F3204C1

About this manual

The document details an analytical method for measuring total sulfur levels and identifying specific sulfur species in motor fuels. It addresses the need for sensitive detection techniques as environmental regulations lower permissible sulfur limits in gasoline.

The study utilizes an Agilent 6890 gas chromatograph coupled with an Agilent 7500a ICP-MS via a GC-ICP-MS interface. The method employs compound-independent calibration using thiophene standards to quantify total sulfur and speciate individual compounds like dimethylbenzothiophenes.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Instrumentation
  • Results and Discussion
  • Sensitivity
  • Compound Independent Calibration
  • Advantages of ICP-MS

Specifications

Gas Chromatograph ModelAgilent 6890 GC
ICP-MS ModelAgilent 7500a ICP-MS
Interface ModelG3158A
Column Dimensions30 M x 0.25 mm id x 0.25 µm
Typical Run Time12 minutes

Frequently asked questions

What is the detection limit for total sulfur in gasoline using this method?

The detection limit (DL) for total sulfur is approximately 0.1 ppm to 0.5 ppm.

Which compounds were used for calibration?

Calibration was based on thiophene and 2-methylthiophene spiked into a solvent mixture.