Agilent Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol In Crude Oil by ASTM Method D7059 Manual

A new method for the analysis of trace methanol in crude oil samples was developed based on two-dimensional heart-cutting gas chromatography. The method was tested on an Agilent 6890N GC system equipped with a new gas phase microfluidic Deans switch that improved the performance of heart-cutting two-dimensional gas chromatography. The results obtained with this method exceeded the performance requirements of ASTM D7059 for the analysis of methanol in crude oil.


Agilent Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol In Crude Oil by ASTM Method D7059 Manual - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
678 KB
Published
02 July, 2012
Updated
01 October, 2023
Model
6890N
MD5 checksum
A3ABF059713828C938BEC047E6EE35D9

About this manual

The document details a new gas phase micro-fluidic Deans switch developed for the Agilent 6890N GC system to improve the performance of heart-cutting two-dimensional gas chromatography. This integrated device replaces traditional plumbing assemblies to reduce thermal mass, minimize leaks, and prevent solvent adsorption, thereby enhancing sensitivity for trace analysis.

What's inside

  • Abstract
  • Introduction
  • Hardware Configuration
  • Instrument Conditions
  • Experimental Procedure
  • Results and Discussion
  • Calibration
  • Quality Control
  • Backflushing
  • Maintenance

Specifications

Secondary ColumnCP-Lowox column, 10 m x 0.53-mm id
Injection Port Temperature325 °C
Oven Final Temperature300 °C
Calibration Range5 to 1000 ppm (m/m)

Frequently asked questions

What is the detection limit requirement for methanol in crude oil according to ASTM Method D7059?

The method requires a signal-to-noise ratio of 5:1 for a 2-ppm standard.

How often should the inlet liner be replaced when analyzing crude oil samples?

It is recommended to change the liner after 50 injections.

When should the inlet body be cleaned during crude oil analysis?

The inlet body may need to be cleaned after 100 injections depending on the samples.