Agilent 6890N Two-Dimensional Gas Chromatographic Analysis of Trace Benzene in Styrene Application Note

A two-dimensional gas chromatographic method (GC-GC) is described for the analysis of trace benzene in styrene. This method uses the Agilent 6890N GC system equipped with a Deans switch device. The method can be used in a prefractionator mode to analyze trace benzene in styrene according to the proposed ASTM method [1]. It can also be used in a heart-cutting mode to transfer more precisely just the benzene peak from the primary column to the sec- ondary column. The Agilent 6890N electronic pneumatics control and flow calculation software are used for setting up the precise flows and pressures required. Excellent calibration and sensitivity of this system is demonstrated in both prefractionator mode and heart-cutting mode, and the interference can be greatly reduced by the heart-cutting.


Agilent 6890N Two-Dimensional Gas Chromatographic Analysis of Trace Benzene in Styrene Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
917 KB
Published
02 July, 2012
Updated
30 September, 2023
Model
6890N
MD5 checksum
516C9CDDBA2E1491809E10FA1502F11A

About this manual

The document details a method for analyzing trace benzene in styrene using the Agilent 6890N GC system equipped with a Deans switch device. The method can be operated in two modes: prefractionator mode to analyze trace benzene according to the proposed ASTM method, and heart-cutting mode to transfer the benzene peak more precisely from the primary column to the secondary column. This approach reduces interference from other hydrocarbons.

What's inside

  • Introduction
  • Deans Switch System Design
  • Heart-Cutting Mode
  • Prefractional Mode
  • Method Development Tool
  • Hardware Configuration
  • Instrument Conditions
  • Results and Discussion
  • Sensitivity
  • Precision

Specifications

Primary columnHP-1 15 m, 0.53 mm, 3.0 ym
Secondary columnINNOWax 60 m, 0.53 mm, 1.0 um
Data systemAgilent multitechnique ChemStation rev. A.08.01

Frequently asked questions

What is the calibration range for benzene in styrene using this method?

The calibration range covers benzene concentrations from 1 ppm to 1000 ppm.

Which software tool is used to set up the Deans switch parameters?

The Deans Switch Calculator is used to determine the correct column flows, inlet pressure, PCM pressure, and restrictor size.