Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application Note

A rapid simulated distillation method is described for analysis of petroleum fractions boiling within the range of ASTM D2887. The method employs an Agilent 6890N gas chromatograph equipped with a split/splitless inlet, flame ionization detector (FID), oven insert, and 3-meter 180-µm id capillary column. Compared to conventional ASTM D2887, the accelerated method yields run times at least two times faster. It yields results comparable to traditional D2887, and agrees with the guidelines of ASTM method D2887 based on published results for the reference gas oil (RGO).


Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
510 KB
Published
02 July, 2012
Updated
28 September, 2023
Model
6890N
MD5 checksum
0AC4734C2C5C43ACE05ED9C030FD7026

About this manual

The document details a rapid simulated distillation method designed for analyzing petroleum fractions boiling within the range of ASTM D2887. The procedure utilizes an Agilent 6890N gas chromatograph equipped with a split/splitless inlet, flame ionization detector (FID), and a short 3-meter capillary column. A SIMDIS software package is integrated into the GC ChemStation to perform data analysis and SimDis calculations.

Compared to conventional ASTM D2887 methods, this accelerated approach yields run times at least two times faster while producing results comparable to traditional standards. The method employs a 3-meter 180-µm id capillary column (DB-1) with specific oven programs and flow rates to achieve high throughput without sacrificing accuracy for reference gas oil analysis.

What's inside

  • Abstract
  • Introduction
  • Experimental Conditions
  • SIMDIS Software
  • Calibration
  • RGO Analysis
  • Cat Cracker Feed Analysis
  • Conclusions

Specifications

Column dimensions3 m x 180 µm x 0.4 µm
Inlet temperature325 °C
Split ratio50:1
FID temperature350 °C
Oven program range35 °C to 340 °C at 30 °C/min
Data acquisition rate5 Hz

Frequently asked questions

What is the speed gain of this method compared to conventional D2887?

The accelerated method yields run times at least two times faster than conventional ASTM D2887.

Does this method agree with ASTM D2887 guidelines?

Yes, it yields results comparable to traditional D2887 and agrees with the guidelines based on published results for reference gas oil.