Agilent 6820 GC System Application Note for Determination of Benzene and Toluene in Gasoline Using ASTM D3606

This application describes a method for determining the content of benzene and toluene in motor and aviation gasoline using an Agilent 6820 gas chromatograph (GC) system and ASTM Method D3606. The method uses an internal standard to calculate the concentration of each component and employs Agilent Cerity NDS software for GC control and data analysis. This application demonstrates that calibration and precision meet or exceed the ASTM D3606 specifications.


Agilent 6820 GC System Application Note for Determination of Benzene and Toluene in Gasoline Using ASTM D3606 - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
65 KB
Published
02 July, 2012
Updated
01 October, 2023
Model
6820
MD5 checksum
A08E689C7EE2656B89B31256BCD3003D

About this manual

This document details an application method for analyzing benzene and toluene in finished motor and aviation gasoline using the Agilent 6820 gas chromatographic (GC) system. The method utilizes a flame ionization detector (FID) and a 6-port valve configuration with OV-101 and TCEP columns, differing from the standard ASTM D3606 method which typically uses a thermal conductivity detector.

The note provides specific experimental configurations, including hardware components like the G1180A GC, 152 packed inlet, and 215 FID. It outlines the method description involving a two-column series setup with a switching valve to separate lighter components from heavier ones. Instrument conditions, such as inlet and detector temperatures set to 250 °C, and flow rates are specified. The document also covers calibration procedures using methyl ethyl ketone as an internal standard and presents results demonstrating that the method meets or exceeds ASTM D3606 specifications for linearity and repeatability.

What's inside

  • Introduction
  • Experimental Configuration
  • Method Description
  • Instrument Conditions
  • Calibration
  • Results and Discussion
  • Sample Injection
  • Conclusions

Specifications

Inlet Temperature250 °C
Detector Temperature250 °C
Oven Temperature90°C
Valve Temperature80°C
TCEP Flow40 mL/min
OV-101 Flow40 mL/min
Air Flow350 mL/min
Hydrogen Flow40 mL/min

Frequently asked questions

Which detector is used in this application method?

The method uses a flame ionization detector (FID), which is noted as easier to maintain than the thermal conductivity detector (TCD) used in the standard ASTM D3606.

What software is used for control and data analysis?

Agilent Cerity NDS software is used for 6820 GC control and data analysis.

What is the correlation coefficient for the calibration curve?

The correlation (R²) of both benzene and toluene exceeds 0.999.