Agilent 1200 Series RRLC Analysis of Phenolic Antioxidant and Erucamide Slip Additives in Polymer

This application note describes the use of the Agilent 1200 Series Rapid Resolution LC (RRLC) system for the separation of antioxidants and erucamide. The system can operate significantly faster than conventional HPLC systems without sacrificing resolution, precision, or sensitivity. The column chemistry and temperature influence on the separation and the sample preparation method are also discussed.


Agilent 1200 Series RRLC Analysis of Phenolic Antioxidant and Erucamide Slip Additives in Polymer - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
283 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
FD6524F51D26A3B9801EDFD1AB95E40E

About this manual

This document details an application for analyzing additives in polymer formulations, specifically phenolic antioxidants and erucamide slip additives. It demonstrates the use of the Agilent 1200 Series Rapid Resolution LC (RRLC) system to separate these components significantly faster than conventional HPLC methods without sacrificing resolution or sensitivity.

The text outlines specific system configurations, including the G1379B microvacuum degasser, G1312B binary pump, G1367B autosampler, G1316B thermostatted column compartment, and G1315C UV/VIS diode array detector. It discusses method optimization using the ZORBAX Eclipse XDB-C18 and ZORBAX SB-C18 columns, mobile phase gradients, and temperature effects on separation speed.

What's inside

  • Introduction
  • System Configuration
  • Experimental Results
  • Fast Method Conversion
  • Optimized Column Temperature
  • Repeatability
  • Real Sample Analysis

Specifications

System ComponentsG1379B microvacuum degasser, G1312B binary pump, G1367B autosampler, G1316B thermostatted column compartment, G1315C UV/VIS diode array detector
Column ChemistryZORBAX Eclipse XDB-C18, ZORBAX SB-C18
Mobile PhaseA: water, B: acetonitrile (ACN)
Analysis Time (Speed-optimized)1.7 minutes
Maximum Column Temperature90 °C

Frequently asked questions

How much faster is the RRLC system compared to conventional HPLC?

The application can achieve comparable results in as few as 3 minutes, which is significantly faster than conventional HPLC methods that often require more than 30 minutes.

What is the minimum detection limit for additives using this method?

The method can determine 1 ppm of additives with very good signal-to-noise response, which exceeds the ASTM specification of 2 ppm.