Agilent 7000 Triple Quadrupole GC/MS Estradiol Detection Application Brief

This application brief describes a highly repeatable method for the quantification of estradiol in human plasma using the Agilent 7000 Triple Quadrupole GC/MS. The method provides an LOD of 1 pg/mL (2 fg on-column) and RSDs for quantification of <15% in the 1 to 600 pg/mL concentration range.


Agilent 7000 Triple Quadrupole GC/MS Estradiol Detection Application Brief - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
166 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
24A4CC5DFC6AEA83C9E4A76F7AC84239

About this manual

This document details a method developed on the Agilent 7890GC/7000 Triple Quadrupole GC/MS system to detect and quantify estradiol in human plasma. The method achieves a limit of detection (LOD) of 1 pg/mL (2 fg on-column) and maintains a relative standard deviation (RSD) for quantification of less than 15% across a concentration range of 1 to 600 pg/mL. The analysis utilizes the MS/MS-NCI (negative chemical ionization) mode with a specific column and injection sequence.

What's inside

  • Abstract
  • Introduction
  • Standards and Reagents
  • Gas Chromatograph and Mass Spectrometer Conditions
  • Results and Discussion
  • Sensitivity and Quantification
  • Peak Area Ratio Precision
  • Conclusion

Specifications

Limit of Detection (LOD)1 pg/mL (2 fg on-column)
Quantification Range1 to 600 pg/mL
Relative Standard Deviation (RSD)<15%
Analytical Column15m x 0.25 mm x 0.25 pm Agilent J&W DB-5MS UI
Carrier GasHelium, constant flow, 1.2 mL/min
MS ModeNCI mode; 40% ammonia
Source Temperature150 °C
Quadrupoles Temperature150 °C
Collision Gas FlowsNitrogen at 1.5 mL/min, Helium at 2.35 mL/min
Total Run Time2.85 minutes

Frequently asked questions

What is the limit of detection for estradiol using this method?

The limit of detection is 1 pg/mL, which corresponds to 2 fg injected on-column.

What is the quantification range for estradiol?

The method provides quantification for estradiol in the range of 1 to 600 pg/mL.

What is the maximum relative standard deviation (RSD) for quantification?

The RSD for quantification never exceeds 15%.

Which ionization mode is used for the analysis?

The instrument is operated in the MS/MS-NCI (negative chemical ionization) mode.