Agilent UHPLC Method Development for Vitamin D2/D3 Separation Application Note

This application note describes the method development options for separation of vitamin D2 and D3 by UHPLC. Partial separation of vitamin D2 and D3 was achieved by an isocratic LC method using a C18 column and methanol/acetonitrile mobile phase. Several optimization parameters were examined to increase resolution (Rs) including column selectivity and column length. A 2.1 mm × 150 mm Agilent ZORBAX StableBond-C18 column provided the best resolution. Further fine tuning included faster flow rates and mobile phase strength resulting in short analysis times of 3 to 1.5 minutes with Rs of 1.6 to 1.4. The performance of a 3.0 mm internal diameter (id) column was also compared to that of the 2.1 mm id column demonstrating higher sample loading before any loss of resolution in the closely spaced peaks.


Agilent UHPLC Method Development for Vitamin D2/D3 Separation Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
545 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
6052C147EBCBBEDF90FC35CE7E69E37A

About this manual

This document describes the optimization of a Liquid Chromatography (LC) method for the separation of Vitamin D2 and Vitamin D3 in food samples. It evaluates various parameters to improve resolution and reduce analysis time, focusing on the transition from standard HPLC to UHPLC methods.

The study compares different column selectivities, specifically between Agilent ZORBAX Eclipse Plus C18 and ZORBAX StableBond-C18 (SB-C18), demonstrating that SB-C18 provides superior resolution for these vitamins. The document further investigates the impact of column length, flow rate, and column internal diameter (ID) on separation efficiency and solvent usage, recommending the use of Agilent RRHD (Robust High Pressure) columns operated at high pressures up to 1200 bar.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Use Column Selectivity to Alter Resolution
  • Use Column Length to Increase N
  • Flow Rate Increase Results
  • 3.0 mm id Solvent Saver Columns
  • Conclusion

Specifications

SystemAgilent 1290 Infinity with G4212A DAD
Detection Wavelength280 nm
Reference Wavelength360 nm
Column Temperature25°C
Column Dimensions2.1 mm x 150 mm

Frequently asked questions

Which column provided the best resolution for Vitamin D2 and D3?

A 2.1 mm x 150 mm Agilent ZORBAX StableBond-C18 provided the best resolution.

What is the maximum operating pressure for the recommended RRHD columns?

The RRHD columns are designed to operate up to 1200 bar.

How does increasing the flow rate affect the analysis time?

Increasing the flow rate from 1 to 1.5 mL/min reduced peak retention by one-third.