Agilent 1290 Infinity LC Application Note for Ultrafast Antioxidant Analysis

The document describes a fast method for the analysis of synthetic antioxidants in vegetable oils. The method uses the Agilent 1290 Infinity LC system equipped with an Agilent ZORBAX Rapid Resolution High Definition (RRHD) column and achieves high throughput in 2 minutes with a backpressure of 1120 bar, which is below the 1200 bar upper limit of the column. Optimization of the mobile phase composition and the temperature are discussed. The figures of merit are illustrated using standard solutions and spiked vegetable oil (sunflower, rapeseed, and olive) extracts. Limits of detection are 1 mg/kg or less in the oil samples. Using a simple methanol extraction, good recovery was obtained for all antioxidants in the oil samples.


Agilent 1290 Infinity LC Application Note for Ultrafast Antioxidant Analysis - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
464 KB
Published
02 July, 2012
Updated
01 October, 2023
MD5 checksum
A22BE194C5AD4D8B068525F44ED63EB0

About this manual

This document details a method for the ultrafast analysis of synthetic antioxidants in edible vegetable oils. The method utilizes the Agilent 1290 Infinity LC system equipped with an Agilent ZORBAX Rapid Resolution High Definition (RRHD) column. By optimizing the mobile phase composition and temperature, the analysis time is reduced to approximately 2 minutes with a backpressure of 1120 bar, which is within the column's upper limit of 1200 bar. The approach allows for high-throughput screening of various antioxidants, including propyl gallate, octyl gallate, dodecyl gallate, and ascorbyl palmitate, in sunflower, rapeseed, and olive oil extracts.

What's inside

  • Abstract
  • Introduction
  • Instrumentation and method
  • Results and Discussion
  • Method performance data
  • Conclusion

Specifications

Analysis time2 min
Backpressure1120 bar
Column upper limit1200 bar
Detection limit1 mg/kg
Column typeZORBAX RRHD Eclipse Plus C18
Column dimensions50 mm L x 2.1 mm id, 1.8 um d
Mobile phase A0.02% phosphoric acid in water
Mobile phase BAcetonitrile/methanol 50/50 or 75/25 v/v

Frequently asked questions

What is the analysis time for the ultrafast method?

The analysis time is reduced to less than 2 minutes.

What is the maximum backpressure of the column used?

The backpressure is 1120 bar, which is below the column's 1200 bar upper limit.

What is the detection limit for antioxidants in oil samples?

The limits of detection are 1 mg/kg or less in the oil samples.

Which vegetable oils were tested in this study?

The study used sunflower, rapeseed, and olive oil extracts.