Reduce Tubing Volume to Optimize Column Performance Application Note

The effects of extra-column volume on 50 and 150 mm columns in 2.1 and 4.6 mm internal diameters were made using a range of capillary tubing volumes installed between the autosampler and column. The results showed that, in using short columns, increases in tubing volume will reduce the column separation efficiency, while increases in tubing volume will have little effect on the separation efficiency of the long column, which is consistent with the results of other scholars.


Reduce Tubing Volume to Optimize Column Performance Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
307 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
8D6FC8D2823398E3F8C4DF6933519191

About this manual

Agilent Technologies presents a study comparing the effects of extra-column volume on 50 and 150 mm HPLC columns with internal diameters of 2.1 mm and 4.6 mm. The research utilized red (0.12 mm id), green (0.17 mm id), and blue (0.25 mm id) capillary tubing installed between the autosampler and column to measure efficiency changes.

The results indicate that decreases in efficiency range from 1% to 31%, depending on the tubing volume and column dimensions. Larger decreases are observed with 2.1-mm ID columns, whereas 4.6-mm ID columns appear nearly immune to extra-column volume effects. The study concludes that minimizing extra-column volume is essential for achieving best results, particularly when using small-volume columns.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Discussion
  • Detector
  • Conclusions

Frequently asked questions

What tubing inside diameter is recommended for small column applications?

The document recommends using capillary tubing with a 0.12 mm inside diameter.

What is the recommended detector cell volume for small volume columns?

A detector cell volume of 2 pL or less is recommended.

How much does efficiency decrease when extra-column volume reaches 7 pL on a 2.1 mm column?

A significant decrease in efficiency is noted at 7 pL.