Agilent 1200 Series Fast Method Development of Salicylic Acid Process Impurities Application Note

This document presents a fast method development for salicylic acid process impurities using Agilent ZORBAX Rapid Resolution High Throughput Columns and Agilent 1200 Series Method Development Solution. The method was developed using ACD/AutoChrom software, allowing for screening of multiple columns, buffers, and organic solvents to find the best experimental conditions. The final result was an isocratic separation achieved in under 3 minutes.


Agilent 1200 Series Fast Method Development of Salicylic Acid Process Impurities Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
444 KB
Published
02 July, 2012
Updated
29 September, 2023
MD5 checksum
776E18CF2DE3B58F1AB2A45A4083AC96

About this manual

The document details a study where salicylic acid and its published process impurities were separated using an Agilent 1200 Series LC system. The method development was automated and controlled by ACD/AutoChrom software version 12.01, which allowed for the screening of multiple columns, buffers, and solvents to quickly identify optimal separation conditions.

What's inside

  • Introduction
  • Experimental
  • Column and Buffer Screening
  • Discussion

Frequently asked questions

What is the maximum pressure the binary pump used in this system can deliver?

The G1312B Binary Pump SL used in the system is capable of delivering up to 600 bar.

At what temperature was the temperature maintained throughout the experiment?

The temperature was maintained at 25 °C throughout the experiment.

What detection wavelength was used to achieve good chromatographic response for the mixture?

Good response for all compounds was found at a 230 nm detection wavelength.