Using Agilent Pesticides Application Kit can quickly and efficiently establish multi-residue pesticide LC/MS/MS methods
The document details the analysis of pesticide residues in food and environmental samples using Liquid Chromatography/Mass Spectrometry/Mass Spectrometry (LC/MS/MS) with a triple quadrupole instrument. It highlights the challenges of monitoring hundreds of analytes and introduces Dynamic Multiple Reaction Monitoring (MRM) as a technique to improve sensitivity and specificity compared to time-segmented methods. The note demonstrates how to set up methods quickly using the Agilent Pesticides Application Kit, which includes a 600-compound pesticide MRM database and pre-tested methods.
| Database Size | 600-compound pesticide MRM database |
|---|---|
| Instrument Technology | Agilent Jet Stream Technology |
| Detection Limit | 100 fg or less |
Dynamic MRM is a technique where analyte ions are monitored only while they are eluting from the LC, significantly improving MS duty cycle time for complex samples.
The kit contains a pesticide test mix, a 600-compound pesticide MRM database, a quick start guide, and several dynamic MRM methods.
The methods were demonstrated using an Agilent 1200 SL Series Rapid Resolution LC and the Agilent 6460 Series Triple Quadrupole LC/MS System.